The near-infrared (NIR) measurement of blood pH relies on the spectral signature of histidine residing on the hemoglobin molecule. If the amount of hemoglobin in solution varies, the size of the histidine signal can vary depending on changes in either the pH or hemoglobin concentration. Multivariate calibration models developed using the NIR spectra collected from blood at a single hemoglobin concentration are shown to predict data from different hemoglobin levels with a bias and slope.