Background: The increased use of ionizing radiation for diagnostic purpose has resulted in an increase in the world population dose. Patient dosimetry in X-ray diagnostic radiology is required to establish diagnostic reference levels (DRLs) and to assess the average dose received by organs and tissues. International bodies have recommended DRLs to be based on dosimetric quantities.
View Article and Find Full Text PDFWithin an International Atomic Energy Agency (IAEA) co-ordinated research project (CRP), a remote end-to-end dosimetric quality audit for intensity modulated radiation therapy (IMRT)/ volumetric arc therapy (VMAT) was developed to verify the radiotherapy chain including imaging, treatment planning and dose delivery. The methodology as well as the results obtained in a multicentre pilot study and national trial runs conducted in close cooperation with dosimetry audit networks (DANs) of IAEA Member States are presented. A solid polystyrene phantom containing a dosimetry insert with an irregular solid water planning target volume (PTV) and organ at risk (OAR) was designed for this audit.
View Article and Find Full Text PDFBackground And Purpose: An audit methodology for verifying the implementation of output factors (OFs) of small fields in treatment planning systems (TPSs) used in radiotherapy was developed and tested through a multinational research group and performed on a national level in five different countries.
Materials And Methods: Centres participating in this study were asked to provide OFs calculated by their TPSs for 10 × 10 cm, 6 × 6 cm, 4 × 4 cm, 3 × 3 cm and 2 × 2 cm field sizes using an SSD of 100 cm. The ratio of these calculated OFs to reference OFs was analysed.