A hybrid quality control (QC) program was developed that integrates automated and conventional Linac QC, realizing the benefits of both automated and conventional QC, increasing efficiency and maintaining independent measurement methods. Failure mode and effects analysis (FMEA) was then applied in order to validate the program prior to clinical implementation. The hybrid QC program consists of automated QC with machine performance check and DailyQA3 array on the TrueBeam Linac, and Delta4 volumetric modulated arc therapy (VMAT) standard plan measurements, alongside conventional monthly QC at a reduced frequency.