The applicability of the ISO slab phantom as calibration phantom for the new ICRU report 95 quantity personal dose should be legitimated by simulations and measurements of backscatter factors on the ISO slab phantom and, comparatively, on a human-like Alderson Rando phantom. An ionization chamber was used to determine backscatter factors for standardized X-ray spectra in the energy range of 16-250 keV and for gamma radiation of 137Cs (662 keV) and 60Co (1250 keV). For the validation of measurement results on the ISO slab, the results were compared with those obtained by means of Monte Carlo simulations using MCNP 6.
View Article and Find Full Text PDFIn this article, the proposal of the International Commission on Radiation Units and Measurements/International Commission on Radiological Protection (ICRU/ICRP), that the ISO slab phantom should continue to be used as a calibration phantom for the new ICRU Report 95 operational quantity personal dose should be legitimised by simulation and performance of experiments to determine backscatter factors on the ISO slab phantom and, in comparison, on an anthropomorphic Alderson Rando phantom. The scope of this work was restricted to the photon energy range of radiation qualities commonly used in x-ray diagnostics. For this purpose, a shadow-free diagnostic (SFD) ionisation chamber was used to measure backscatter factors for X radiation in the energy range of 24 keV to 118 keV.
View Article and Find Full Text PDF