Background: The Nippy Clearway is a new mechanical insufflation-exsufflation device used to assist cough.
Methods: We compared the peak expiratory flow (PEF) with the Nippy Clearway versus the CoughAssist in a bench experiment. The relationship between PEF and pressure at the airway opening during exsufflation (minimum expiratory PAO) was investigated under 6 combinations of compliance (30 or 60 mL/cm H2O) (C30 and C60) and resistance (0, 5, or 20 cm H2O/L/s) (R0, R5, and R20) over a 25-50 cm H2O range of set P(AO).
The patient information and obtaining their consent before any therapeutic constitute a basis for medicine exercise. However, there are situations where the physician may be in default. In that case, concepts of benefit/risk balance, of clear and appropriate information, and of legal liability for medico-surgical staff takes all their magnitude.