Background: Availability of whole blood creatinine estimation for patients scheduled to undergo radiological contrast investigations can provide information to aid patient care by reducing adverse effects and improving departmental efficiencies.
Methods: We performed imprecision studies, different patient sample type comparison in 40 participants, and a limited interference study with dopamine and dobutamine on the i-Stat and StatStrip point-of-care enzymatic analysers with the Beckman DxC800 Jaffe assay.
Results: Imprecision results showed that the i-Stat performed better.