The purpose of this study was to determine the performance characteristics of the Cozart Methadone Microplate ELISA assay for the detection of methadone and methadone metabolites in hair specimens. One hundred and ten hair specimens were collected from volunteers (n=46) with a history of drug use and from drug-related deaths (n=64). The hair samples (approximately 20 mg) were extracted by sonication in methanol followed by overnight extraction in methanol at 60 degrees C.
View Article and Find Full Text PDFThe purpose of this study was to determine the performance characteristics of the Cozart microplate EIA for detecting opiates in oral fluid from patients in a drug misuse treatment program. Oral fluid samples were collected using the Cozart RapiScan Collection System from 216 donors who were receiving treatment for their addiction and were monitored for drug abuse. A further 40 oral fluid samples were collected from volunteer donors who were not drug users.
View Article and Find Full Text PDFThe purpose of this study was to determine the performance characteristics of the Cozart Amphetamine Microplate EIA for detecting amphetamine in oral fluid. Oral fluid samples were collected using the Cozart RapiScan Collection System from 135 volunteer donors from drug treatment clinics. A further 35 oral fluid samples were collected from volunteer donors who were not drug users.
View Article and Find Full Text PDFThe purpose of these studies was to evaluate the performance characteristics of the Cozart Microplate Enzyme Immunoassay (EIA) for the determination of methadone in oral fluid from patients in a drug misuse treatment program. Oral fluid specimens were collected using the Cozart RapiScan Collection system from 198 donors who were receiving treatment for their addiction and were monitored for drug misuse. Oral fluid specimens were also collected from forty volunteer donors who were not drug users.
View Article and Find Full Text PDFThe objective of this study was to investigate the accuracy of screening postmortem whole blood for oxycodone using the ratio of the oxycodone immunoassay response to the response for the specimen obtained with a general opiate-class immunoassay. Fifty eight specimens which were negative for opiates and 158 postmortem whole blood specimens positive for opiates including 66 specimens known to contain oxycodone were assayed. Specimens were diluted 1:5 with assay buffer and analyzed by both the Neogen Oxymorphone/Oxycodone ELISA and the Neogen Opiate Group ELISA (Neogen Corporation, Lexington KY).