A thermoluminescence dosimetry (TLD) system was characterised at SSDL-ININ to verify the air-kerma strength (S(K)) and dose-to-water (D(W)) values for (137)Cs sources used in low dose rate (LDR) brachytherapy treatments at the Hospital General de Mexico (HGM). It consists of a Harshaw 3500 reader and a set of TLD-100 powder capsules. The samples of TLD-100 powder were calibrated in terms of D(W) vs.
View Article and Find Full Text PDFMeasurements of absorbed dose at 5 cm depth in a 30 x 30 x 30 cm3 water phantom have been performed using three independent dosimetric techniques: Fricke, alanine and radiochromic dye film (GafChromic HD-810). The measurements were carried out in the secondary standard dosimetry laboratory at ININ Mexico using a collimated 60Co gamma source with a radiation field of 10 x 10 cm2 at the phantom front surface. The source to phantom distance was set at 100 cm.