Radiol Bras
January 2023
Objective: To develop an automated co-registration system and test its performance, with and without a fiducial marker, on single-photon emission computed tomography (SPECT) images.
Materials And Methods: Three SPECT/CT scans were acquired for each rotation of a Jaszczak phantom (to 0°, 5°, and 10° in relation to the bed axis), with and without a fiducial marker. Two rigid co-registration software packages-SPM12 and NMDose-coreg-were employed, and the percent root mean square error (%RMSE) was calculated in order to assess the quality of the co-registrations.
Objective: The purpose of this study was to evaluate how statistical fluctuation in single-photon emission computed tomography (SPECT) images propagate to absorbed dose maps.
Methods: SPECT/computed tomography (CT) images of iodine-131 filled phantoms, using different acquisition and processing protocols, were evaluated using STRATOS software to assess the absorbed dose distribution at the voxel level. Absorbed dose values and coefficient of variation (COV) were analyzed for dosimetry based on single time-point SPECT images and time-integrated activities of SPECT sequences with low and high counts.