Treatment of cancer is carried out using photon beams from high-energy medical linear accelerators. Photo-neutrons are also produced as an unwanted by product in the process of dose delivery to the cancer patients during their radiation treatments. In the present study, photo-neutron dose equivalents (both thermal and fast components) per unit delivered gamma-photon dose were measured at different depths, as function of distances from iso-centre in patient plane, field sizes, wedge angles and at LINAC head for a 15-MV medical linear accelerator model Elekta Precise using multi-foil activation technique.
View Article and Find Full Text PDF