Purpose/objectives: (A) To examine the alignment accuracy of CBCT guidance for brain metastases with off centered isocenters, (B) to test dose delivery and targeting accuracy for single isocenter treatments with multiple brain metastases. We report the results of the end-to-end test for Truebeam stereotactic radiosurgery (SRS).
Materials/methods: An anthropomorphic CT head phantom was drilled with five MOSFET inserts and two PTW Pinpoint chamber inserts.
Daily output variations of up to ±2% were observed for a protracted time on a Varian TrueBeam® STx; these output variations were hypothesized to be the result of atmospheric communication of the sealed monitor chamber. Daily changes in output relative to baseline, measured with an ionization chamber array (DQA3) and the amorphous silicon flat panel detector (IDU) on the TrueBeam®, were compared with daily temperature-pressure corrections (P ) determined from sensors within the DQA3. Output measurements were performed using a Farmer® ionization chamber over a 5-hour period, during which there was controlled variation in the monitor chamber temperature.
View Article and Find Full Text PDFThis work quantified differences between recommendations of the TG-51 and TG-51 addendum reference dosimetry protocols. Reference dosimetry was performed for flattened photon beams with nominal energies of 6, 10, 15, and 23 MV, as well as flattening-filter free (FFF) beam energies of 6 and 10 MV, following the recommendations of both the TG-51 and TG-51 addendum protocols using both a Farmer ionization chamber and a scanning ionization chamber with calibration coefficients traceable to absorbed dose-to-water (D ) standards. Differences in D determined by the two protocols were 0.
View Article and Find Full Text PDFPurpose: Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry.
View Article and Find Full Text PDFPurpose: Gafchromic(®) EBT2 film has a yellow marker dye incorporated into the active layer of the film that can be used to correct the film response for small variations in thickness. This work characterizes the effect of the marker-dye correction on the uniformity and uncertainty of dose measurements with EBT2 film. The effect of variations in time postexposure on the uniformity of EBT2 is also investigated.
View Article and Find Full Text PDF