Purpose: The purpose of this study was to provide guidance on quality management for electronic brachytherapy.
Materials And Methods: The task group used the risk-assessment approach of Task Group 100 of the American Association of Physicists in Medicine. Because the quality management program for a device is intimately tied to the procedure in which it is used, the task group first designed quality interventions for intracavitary brachytherapy for both commercial electronic brachytherapy units in the setting of accelerated partial-breast irradiation.
Purpose: In this study, newly formulated XR-RV3 GafChromic film was calibrated with National Institute of Standards and Technology (NIST) traceability for measurement of patient skin dose during fluoroscopically guided interventional procedures.
Methods: The film was calibrated free-in-air to air kerma levels between 15 and 1100 cGy using four moderately filtered x-ray beam qualities (60, 80, 100, and 120 kVp). The calibration films were scanned with a commercial flatbed document scanner.