Objectives: Employ Micro-X-ray diffraction and temperature-modulated differential scanning calorimetry to investigate microstructural phases, phase transformations, and effects of heat treatment for rotary nickel-titanium instruments.
Methods: Representative as-received and clinically used ProFile GT and ProTaper instruments were principally studied. Micro-XRD analyses (Cu Kalpha X-rays) were performed at 25 degrees C on areas of approximately 50 microm diameter near the tip and up to 9 mm from the tip.
Dental schools across America are challenged by the shortage of qualified faculty to fill vacant positions. This project, conducted through survey methodology, focused on obtaining a basic understanding of the personality types that seek out and maintain positions as full-time endodontic educators and compared those educators with endodontic residents to define similarities and differences that could lead to strategies to recruit those residents into academic endodontic education to assist in alleviating the faculty shortage in dentistry. All full-time endodontic faculty and residents were invited to participate.
View Article and Find Full Text PDFA contributing mechanism for clinical failure of NiTi rotary instruments might be excessive work hardening. Vickers hardness was measured with 300 gm load near the flutes in three regions (D2-D4, D6-D10, and D14 towards the shank) for nine representative clinically used ProFile GT instruments that had been axially sectioned (10 measurements in each region). Consistent values could not be obtained at D1.
View Article and Find Full Text PDFThe aim of this study was to determine the remaining dentin thickness (RDT) in the apical 4 mm following four cleaning and shaping techniques. Sixty human adult extracted mandibular incisors and 60 mesial buccal canals of mandibular molars were assigned to five groups of 12 teeth for each tooth type: Step-down stainless steel hand instrumentation, Lightspeed, Profile GT and 0.4 taper, K3 "g pack," control group.
View Article and Find Full Text PDFNumerous discarded ProFile GT, ProFile, and ProTaper nickel-titanium rotary instruments obtained from two graduate endodontic clinics were examined by scanning electron microscopy. These instruments had an unknown history of clinical use and had fractured or experienced considerable permanent torsional deformation without complete separation. The failure processes generally exhibited substantial ductile character, evidenced by a dimpled rupture fracture surface.
View Article and Find Full Text PDFDiscarded ProFile and ProTaper nickel-titanium rotary instruments, with unknown history of clinical use, were obtained from graduate endodontic clinics at Ohio State University and University of Texas Health Science Center at Houston Dental Branch. These discarded instruments and as-received instruments of both types were examined with a scanning electron microscope to investigate effects of clinical use and causes of failure. For used ProTaper instruments, dentinal debris was wedged mostly in narrow, radial, land-type regions and less on convex flute surfaces.
View Article and Find Full Text PDFThe appearances of the tip sections of ProFile 0.04 taper and Lightspeed 25-mm long, ISO size 25, nickel-titanium rotary instruments were compared with a scanning electron microscope in the as-received condition and after one, three, and six simulated clinical uses to prepare mesial canals of extracted mandibular molars. For the used ProFile instruments, there was some flattening of the characteristic material rollover and minor apparent wear at the edges of the flutes, but there was little change in the tip regions of the used Lightspeed instruments.
View Article and Find Full Text PDFIt has been shown that rotary nickel-titanium files cannot be used indefinitely. Researchers and clinicians have not been able to agree on how many times a file can be used before being discarded, except if a file has fractured or become visibly distorted. This study used ISO size 20 files of 0.
View Article and Find Full Text PDF