Measurement of the acoustic power in high-energy ultrasonic devices is complex due to occurrence of the strong cavitation in front of the sonotrode tip. In our research we used three methods for characterization of our new ultrasonic probe for neuroendoscopic procedures. The first method is based on the electromechanical characterization of the device measuring the displacement of the sonotrode tip and input electrical impedance around excitation frequency with different amounts of the applied electrical power The second method is based on measuring the spatial pressure magnitude distribution of an ultrasound surgical device produced in an anechoic tank.
View Article and Find Full Text PDFIn this paper, our experience with originally constructed Neurosurgical Endoscopic Contact Ultrasound Probe "NECUP-2" in neuroendoscopy is reported. Between June 1997 and June 2007, 132 neuroendoscopic procedures have been performed: 102 endoscopic thrid ventriculostomies (ETV), 15 arachnoid cysts and 5 intraventricular tumours operations. The "NECUP-2" was applied effectively in all cases in which blunt perforation was not possible: 38/102 ETY, 10/10 septostomies, 15/15 arachnoid cysts.