Exposure to spark discharges may occur beneath high voltage transmission lines when contact is initiated with a conductive object (such as a motor vehicle) with the spark discharge mediated by the ambient electric field from the line. The objective of this study was to assess whether such exposures could interfere with the normal functioning of implanted cardiac pacemakers (PMs). The experiment consisted of PMs implanted in a human-sized phantom and then exposed to spark discharge through an upper extremity.
View Article and Find Full Text PDFElectric and magnetic field exposure limits published by International Commission for Non-Ionizing Radiation Protection and Institute of Electrical and Electronics Engineers are aimed at protection against adverse electrostimulation, which may occur by direct coupling to excitable tissue and, in the case of electric fields, through indirect means associated with surface charge effects (e.g. hair vibration, skin sensations), spark discharge and contact current.
View Article and Find Full Text PDFFor magnetic field exposures at extremely low frequencies, the electrostimulatory response with the lowest threshold is the magnetophosphene, a response that corresponds to an adult exposed to a 20 Hz magnetic field of nominally 8.14 mT. In the IEEE standard C95.