Publications by authors named "Tang-Chin Kuan"

We evaluated the Wellion Galileo GLU/KET blood and ketone (ß-Hydroxybutyrate, ß-OHB) meter to demonstrate that it meets ISO15107:2015 regulatory approval criteria. A total of 100 subjects (52 female, age: 30 to 84 years, diabetes: 10 type 1/90 type 2) with blood glucose levels distributed over the entire measurement range as required by the ISO15197 protocol were tested (double determinations with 3 strip lots and two devices). A similar test protocol was followed to test ß-OHB strip performance (reference devices: YSI 2300plus for glucose and STANBIO ß-HOB LiquiColor TestKit for ß-OHB).

View Article and Find Full Text PDF