A polystyrene (PS)-based composite resin dosimeter (CRD) film comprising 2-(phenylamino)-6-(dipentylamino)-3-methylspiro[9-xanthene-9,3'-phthalide] (Black305) fluoran leuco dye, cerium-doped yttrium aluminum perovskite YAlO:Ce (YAP:Ce) scintillator, and 2-(4-methoxystyryl)-4,6-bis(trichloromethyl)-1,3,5-triazine (MBTT) photoacid generator was developed for the visualization of X-ray exposure doses in the range of 18-170 Gy. The color of the CRD film changed from yellow to black within this range based on the X-ray exposure dose. A CRD film based on Black305/YAP:Ce/MBTT/PS could be used to produce a dosimeter with various color change sensitivities by controlling the thermodynamic parameters of the constituent materials.
View Article and Find Full Text PDFACS Appl Mater Interfaces
April 2018
Polystyrene (PS)-based composite microfibers combined with a photochromic spiropyran dye, 1,3,3-trimethylindolino-6'-nitrobenzopyrylospiran (6-nitro BIPS), and a photostimulable phosphor, europium-doped barium fluorochloride (BaFCl:Eu), were developed for the detection of X-ray exposure doses on the order of approximately 1 Gy. To produce the PS-based composite microfibers, we employed a forcespinning method that embeds a high concentration of phosphor in PS in a safe, inexpensive, and simple procedure. On the basis of the optimization of the forcespinning process, fibrous color dosimeters with a high radiation dose sensitivity of 1.
View Article and Find Full Text PDF