Background: The widely automated method using indirect ion specific electrodes (ISE) potentiometry for determination of sodium concentration is prone to interference from lipaemia. Manufacturer-specified lipaemic (L)-index cut offs may underestimate the effects of endogenous lipaemia.
Methods: We assessed the interference on sodium concentration caused by endogenous lipaemia in 32 residual samples (from 13 patients) using indirect ISE (Cobas 8000 modular analyser with c702 module, Roche diagnostics) and direct ISE (GEM 4000 premier, Werfen) potentiometric methods.
Background: The target values for plasma glucose concentrations for the investigation and diagnosis of diabetes and impaired fasting glucose, and the realization that small incremental changes in glucose concentration increase the risk of adverse events, has led to greater focus on laboratory glucose results. Although analytical methods show acceptable precision, the control of preanalytical error due to the stability of glucose remains problematic. The aim of this study was to compare glucose concentrations in 3 different and commercially available blood tubes, with analysis and storage under current practices and conditions.