The present study aimed to determine the effect of femoral component materials and sterilization methods on wear properties of total knee prostheses by using a knee simulator test and retrieval analysis. The simulator test revealed that ultrahigh molecular weight polyethylene (UHMWPE) inserts had remarkably lower wear against the ceramic femoral component than against the Co-Cr femoral component. However, the retrieval study revealed no significant difference in the linear wear between the former and the latter.
View Article and Find Full Text PDFClin Orthop Relat Res
February 2008
Unlabelled: To augment cement-bone bonding, we interposed hydroxyapatite granules at the cement-bone interface (bioactive bone cement technique). Hydroxyapatite granules (2-3 g) were smeared on the bone surface of the acetabulum and femur just before cementing. We used porous hydroxyapatite granules 300 to 500 microm in diameter from 1986 to 1988 (first generation) and granules 100 to 300 microm in diameter from 1989 to 1991 (second generation).
View Article and Find Full Text PDFClin Orthop Relat Res
February 2008
Some previous studies suggest that aging influences wear and oxidatively degraded nonsterilized ultra-high-molecular-weight polyethylene (UHMWPE) exhibits decreased wear resistance. We therefore asked whether shelf-aging storage conditions influenced degradation and wear resistance of gamma-irradiated UHMWPE. We examined oxidation and wear of 100-Mrad gamma-irradiated UHMWPE (100-Mrad polyethylene) cups shelf-aged for 30 years without (n=2) or with (n=2) packages.
View Article and Find Full Text PDFTo augment cement-bone bonding in total hip arthroplasty, we interposed hydroxyapatite (HA) granules at the interface. We report the histologic findings of 2 cases 14 and 18 years after cemented total hip arthroplasty with this technique. We found that a thick bony layer was observed where rich layers of HA granules existed.
View Article and Find Full Text PDFStudies have shown that gamma-irradiation of polyethylene (PE) generally results in degradation by surface oxidation. However, from 1970 to 1978 Oonishi et al. used ultra-high-molecular-weight polyethylene (UHMWPE) cross-linked and sterilized by 100 Mrad of gamma-irradiation in air (100 Mrad PE) for total hip prostheses, and obtained excellent clinical results extending for 30 years.
View Article and Find Full Text PDFThe wear characteristics of ultra-high-molecular-weight polyethylene acetabular cups irradiated with 6.0 to 7.5 Gy of gamma-irradiation were studied in a hip simulator and in vivo in 70 hips for 6 to 39 months.
View Article and Find Full Text PDFJ Biomed Mater Res B Appl Biomater
August 2005
The alumina-ceramic total knee prosthesis developed by Kyocera Corp. was implanted in 1979, and was in clinical use for 23 years until total knee arthroplasty revision surgery in January 2002. It is believed that this is the longest clinical period of a ceramic total knee prosthesis reported to date in the world.
View Article and Find Full Text PDF