Arrhythmia-induced beat-to-beat variability of blood pressure (BP) is deemed to hinder the reliability of non-invasive oscillometric measurements (NIBP) but few data support this belief. We assessed the impact of arrhythmia on a NIBP device never tested for this purpose. We compared, in intensive care unit patients with and without arrhythmia, the agreement between three pairs of NIBP (Infinity™ Delta monitor, Dräger medical systems) and invasive readings.
View Article and Find Full Text PDFBackground: Continuous and intermittent noninvasive measurements of arterial blood pressure (BP) have not been compared in the same population. In a large panel of intensive care unit patients, we assessed the agreement between CNAP™ (Continuous Noninvasive Arterial Pressure) finger cuff beat-to-beat monitoring of BP and reference intraarterial measurements. Two automated oscillometric brachial cuff devices were also tested: CNAP brachial cuff (used for CNAP finger cuff calibration) and an alternative device.