Introduction: Current dosimetry protocols for clinical protons using air-filled ionization chambers assume that the perturbation correction factor is equal to unity for all ionization chambers and proton energies. Since previous Monte Carlo based studies suggest that perturbation correction factors might be significantly different from unity this study aims to determine perturbation correction factors for six plane-parallel and four cylindrical ionization chambers in proton beams at clinical energies.
Materials And Methods: The dose deposited in the air cavity of the ionization chambers was calculated with the help of the Monte Carlo code TOPAS/Geant4 while specific constructive details of the chambers were removed step by step.
To provide Monte Carlo calculated beam quality correction factors (k ) for monoenergetic proton beams using [Formula: see text], a toolkit based on the Monte Carlo code [Formula: see text]. Monte Carlo simulations of six plane-parallel and four cylindrical ionization chambers were carried out. The latest ICRU 90 recommendations on key data for ionizing-radiation dosimetry were used to calculate the electronic stopping powers and to select the mean energy necessary to create an ion pair in air ([Formula: see text]).
View Article and Find Full Text PDF