We developed a system to adjust the rate of a continuous rocuronium (Rb) infusion to achieve 3 ≤ %T1 ≤ 10 with a closed-loop control. Samples were collected from 15 patients, and Rb blood concentrations were measured at the following time points: (1) when %T1 recovered to 3% or more after the initial Rb infusion; (2) when %T1 stabilized within the target range; (3) at the cessation of the Rb infusion; (4) 5 min after the sugammadex administration. The predicted Rb blood concentration at each time point was calculated and recorded using the pharmacokinetic parameters of Wierda et al.
View Article and Find Full Text PDFBackground: We previously developed an automated total intravenous anesthesia control system that uses new closed-loop system algorithms to administer propofol, remifentanil, and rocuronium based on the bispectral index and train-of-four data. We recently improved this automated control system by adding a safety mechanism and using a modified monitoring device.
Methods: Patients scheduled for elective surgery were randomly assigned to closed-loop feedback control (automatic group) or the manual administration of propofol, remifentanil, and rocuronium (manual group).