Purpose: This paper outlines the commissioning of the Varian (VMS, Varian Medical Systems, Palo Alto, CA) Universal Interstitial Cylinder (UIC) applicator set for Ir-192 HDR brachytherapy. The UIC was commissioned for use with CT and MRI and a custom phantom was designed to avoid the introduction of water-like materials into the needle guide tracks. Various marker strands were investigated to determine which allowed the most accurate reconstruction of source positions.
View Article and Find Full Text PDFAm J Ophthalmol Case Rep
December 2024
Purpose: Patients with conjunctival squamous cell carcinoma that present with persisting disease or recurrence following topical chemotherapy and/or surgery especially when invading the sclera are challenging to treat. Herein, we describe the use of high-dose-rate (HDR), FDA-cleared, yttrium-90 (Y) plaque brachytherapy for such lesions.
Observation: Three cases of invasive conjunctival squamous cell carcinoma that had exhibited a poor response or recurrence following topical chemotherapy and/or surgery are described.
Purpose: To design, manufacture, and validate a female pelvic phantom for multi-modality imaging (CT, MRI, US) to benchmark a commercial needle tracking system with application in HDR gynecological (GYN) interstitial procedures.
Materials And Methods: A GYN needle-tracking phantom was designed using CAD software to model an average uterus from a previous patient study, a vaginal canal from speculum dimensions, and a rectum to accommodate a transrectal ultrasound (TRUS) probe. A target volume (CTV ) was designed as an extension from the cervix-uterus complex.
The aim of this work is to describe the implementation and commissioning of a plaque brachytherapy program using Eye Physics eye plaques and Plaque Simulator treatment planning system based on the experience of one institution with an established COMS-based plaque program. Although commissioning recommendations are available in official task groups publications such as TG-129 and TG-221, we found that there was a lack of published experiences with the specific details of such a transition and the practical application of the commissioning guidelines. The specific issues addressed in this paper include discussing the lack of FDA approval of the Eye Physics plaques and Plaque Simulator treatment planning system, the commissioning of the plaques and treatment planning system including considerations of the heterogeneity corrected calculations, and the implementation of a second check using an FDA-approved treatment planning system.
View Article and Find Full Text PDFPurpose: For medically inoperable endometrial cancer (MIEC), the volumetric target of image-guided brachytherapy (IGBT) techniques is not well established. We propose a high-risk CTV (HRCTV) concept and report associated rates of local control and toxicity.
Methods And Materials: For all MIEC patients receiving definitive external beam radiotherapy (EBRT) followed by MRI-based IGBT at a single institution, BT dose was prescribed to HRCTV defined as GTV plus endometrial cavity with a planning goal of a summed EQD2 D90 of ≥85 Gy.
Purpose: Our purpose was to review the preliminary outcomes of patients with uveal melanoma treated with iodine-125 plaques using a novel treatment planning approach
Methods And Materials: This was a single institution, retrospective review of patients treated with iodine-125 brachytherapy for uveal melanoma from November 2016 to February 2019. We used 3-dimensional treatment planning with the Eye Physics Plaque Simulator to ensure that a minimum of 63 Gy covered a 2-mm circumferential tumor margin and the apex height of the tumor over 94 hours. Primary endpoints were local failure, systemic metastasis, final visual acuity (VA), and radiation retinopathy.
The American Association of Physicists in Medicine (AAPM) formed Task Group 178 (TG-178) to perform the following tasks: review in-phantom and in-air calibration protocols for gamma stereotactic radiosurgery (GSR), suggest a dose rate calibration protocol that can be successfully utilized with all gamma stereotactic radiosurgery (GSR) devices, and update quality assurance (QA) protocols in TG-42 (AAPM Report 54, 1995) for static GSR devices. The TG-178 report recommends a GSR dose rate calibration formalism and provides tabulated data to implement it for ionization chambers commonly used in GSR dosimetry. The report also describes routine mechanical, dosimetric, and safety checks for GSR devices, and provides treatment process quality assurance recommendations.
View Article and Find Full Text PDFPurpose: As the utilization of brachytherapy procedures continues to decline in clinics, a need for accessible training tools is required to help bridge the gap between resident comfort in brachytherapy training and clinical practice. To improve the quality of intracavitary and interstitial high-dose-rate brachytherapy education, a multimaterial, modular, three-dimensionally printed pelvic phantom prototype simulating normal and cervical pathological conditions has been developed.
Methods And Materials: Patient anatomy was derived from pelvic CT and MRI scans from 50 representative patients diagnosed with localized cervical cancer.
Background: Treatment planning for I-125 plaque therapy for uveal melanoma has advanced significantly since the Collaborative Ocular Melanoma Study trial, with more widely available image-guided planning and improved dosimetry.
Objective: We evaluated real-world practice patterns for I-125 plaque brachytherapy in the United States by studying practice patterns at centers that comprise the Ocular Oncology Study Consortium (OOSC).
Methods: The OOSC database and responses to a treatment practice survey were evaluated.
Int J Radiat Oncol Biol Phys
April 2016
Purpose: To calculate vaginal doses during image guided brachytherapy with volume-based metrics and correlate with long-term vaginal toxicity.
Methods And Materials: In this institutional review board-approved study, institutional databases were searched to identify women undergoing computed tomography and/or magnetic resonance-guided brachytherapy at the Duke Cancer Center from 2009 to 2015. All insertions were contoured to include the vagina as a 3-dimensional structure.
No accepted official protocol exists for the dosimetry of the Leksell Gamma Knife (GK) stereotactic radiosurgery device. Establishment of a dosimetry protocol has been complicated by the unique partial-hemisphere arrangement of 201 individual 60Co beams simultaneously focused on the treatment volume and by the rigid geometry of the GK unit itself. This article proposes an air kerma based dosimetry protocol using either an in-air or in-acrylic phantom measurement to determine the absorbed dose rate of fields of the 18 mm helmet of a GK unit.
View Article and Find Full Text PDF