January 2014
Applying Monte Carlo method, the present paper simulates the emitted X-ray spectrum of miniature X-ray tube with thirteen thickness of beryllium window in the range from 50 to 500 microm. By analyzing the characteristic of the spectrums, the reasonable choice of thickness of beryllium window relies on the application and for the beryllium window it is not the thinner the better. Taking in-situ EDXRF as an example, though the emission X-ray intensity is higher as the thickness of the beryllium window becomes thinner, the proportion of useless low-energy X-ray (<5 keV) intensity to all energy X-ray intensity also is higher (>20%).
View Article and Find Full Text PDFGuang Pu Xue Yu Guang Pu Fen Xi
November 2013
Thirteen rock samples were collected for studying the variation of element content in the mineral during the alteration process from Xinjiang, China. The IED-6000 in situ micro energy dispersive X-ray fluorescence developed by CDUT was applied to get chemical and physical data from minerals. The non-destructive spectrometer is based on a low-power Mo-anode X-ray tube and a Si-PIN peltier cooled X-ray detector.