Purpose: Clinical investigational studies were conducted to demonstrate the accuracy and reproducibility of the Illumina MiSeqDx CF System, a next-generation sequencing (NGS) in vitro diagnostic device for cystic fibrosis testing.
Methods: Two NGS assays - a Clinical Sequencing Assay (Sequencing Assay) and a 139-Variant Assay (Variant Assay) - were evaluated in both an Accuracy Study and a Reproducibility Study, with comparison to bi-directional Sanger sequencing and PCR as reference methods. For each study, positive agreement (PA), negative agreement (NA), and overall agreement (OA) were evaluated.
Objectives: Fetal mutations and fetal chromosomal abnormalities can be detected by molecular analysis of circulating cell free fetal DNA (ccffDNA) from maternal plasma. This comprehensive study was aimed to investigate and verify blood collection and blood shipping conditions that enable Noninvasive Prenatal Testing. Specifically, the impact of shipping and storage on the stability and concentration of circulating cell-free DNA (ccfDNA) in Streck® Cell-Free DNA™ Blood Collection Tubes (Streck BCTs, Streck, Omaha NE).