Publications by authors named "Seng B Lim"

Background: Advanced bilateral breast cancers with local and regional nodal involvement are rare and challenging to plan due to conflicting demands of conformality, uniformity and sparing of organs at risk such as heart and lungs. Pencil beam scanning (PBS) protons have been shown to provide improved organ at risk (OAR) sparing, conformality and homogeneity compared to photon techniques including three-dimensional conformal radiotherapy, intensity modulated radiation therapy (IMRT), volumetric arc therapy (VMAT), and tomotherapy. We performed a blinded comparison between VMAT and PBS plans and extend this comparison with a case study of a patient win expander during radiotherapy.

View Article and Find Full Text PDF
Article Synopsis
  • * A study evaluated the myQA SRS device, a high-resolution CMOS 2D detector, for its performance in PSQA by characterizing its dosimetric properties and conducting gamma index analyses on nine simulated and clinical plans.
  • * Results indicated that the myQA SRS showed excellent dosimetric agreement with dosimetry films, achieving average gamma index pass rates of 99.9% and 99.2%, demonstrating its reliability and potential as a modern alternative to traditional film dosim
View Article and Find Full Text PDF

Objective: To investigate the feasibility of standardizing RT simulation CT scanner protocols between vendors using target-based image quality (IQ) metrics.

Method And Materials: A systematic assessment process in phantom was developed to standardize clinical scan protocols for scanners from different vendors following these steps: (a) images were acquired by varying CTDI and using an iterative reconstruction (IR) method (IR: iDose and model-based iterative reconstruction [IMR] of CT-Philips Big Bore scanner, SAFIRE of CT-Siemens biograph PETCT scanner), (b) CT exams were classified into body and brain protocols, (c) the rescaled noise power spectrum (NPS) was calculated, (d) quantified the IQ change due to varied CTDI and IR, and (e) matched the IR strength level. IQ metrics included noise and texture from NPS, contrast, and contrast-to-noise ratio (CNR), low contrast detectability (d').

View Article and Find Full Text PDF

Background: Electromagnetic transponders bronchoscopically implanted near the tumor can be used to monitor deep inspiration breath hold (DIBH) for thoracic radiation therapy (RT). The feasibility and safety of this approach require further study.

Methods: We enrolled patients with primary lung cancer or lung metastases.

View Article and Find Full Text PDF

Automated assessment of noise level in clinical computed tomography (CT) images is a crucial technique for evaluating and ensuring the quality of these images. There are various factors that can impact CT image noise, such as statistical noise, electronic noise, structure noise, texture noise, artifact noise, etc. In this study, a method was developed to measure the global noise index (GNI) in clinical CT scans due to the fluctuation of x-ray quanta.

View Article and Find Full Text PDF
Article Synopsis
  • - The study introduces a new calibration method to address signal drift in Respiratory Gating for Scanner systems, specifically when using a wall-mounted camera and dealing with table sagging.
  • - The calibration involved simulating a patient's weight using solid water phantoms and measuring at various points around the CT isocenter in two hospitals.
  • - Results showed that the new method significantly reduced signal drift during scans, enhancing accuracy in radiation delivery for patients, particularly those with higher body weight.
View Article and Find Full Text PDF

Purpose: This study reports clinical experience and feasibility of using a 2-dimensional (2D)-kV image system with online intervention in the ultrafractionated stereotactic body radiation treatment (UF-SBRT) of prostate cancer.

Methods And Materials: Fifteen patients with prostate cancer who had a low- to intermediate-risk marker implanted received UF-SBRT with online 2D-kV image tracking and a manual beam interruption strategy with a 2-mm motion threshold. A total of 180 kV paired setup images and 1272 intrabeam 2D-kV images were analyzed to evaluate the setup deviation and intratreatment target deviation.

View Article and Find Full Text PDF

A new film dosimetry system consists of the new OrthoChromic™ OC-1 film, and a novel calibration procedure was evaluated. Two films, C1 and C2, were exposed simultaneously using the 6FFF beam with a step-wedge pattern of five steps ranging from 590 to 3000 cGy. C1 was used for calibration, and C2 was used for calibration curve validation.

View Article and Find Full Text PDF

Purpose: End-to-end testing (E2E) is a necessary process for assessing the readiness of the stereotactic radiosurgery (SRS) program and annual QA of an SRS system according to the AAPM MPPG 9a. This study investigates the differences between using a new SRS MapCHECK (SRSMC) system and an anthropomorphic phantom film-based system in a large network with different SRS delivery techniques.

Methods And Materials: Three SRS capable Linacs (Varian Medical Systems, Palo Alto, CA) at three different regional sites were chosen to represent a hospital network, a Trilogy with an M120 multi-leaf collimator (MLC), a TrueBeam with an M120 MLC, and a TrueBeam Stx with an HD120 MLC.

View Article and Find Full Text PDF

Purpose: Commercial independent monitor unit (IMU) check systems for high-magnetic-field MR-guided radiation therapy (RT) systems are lacking. We investigated the feasibility of adopting an existing treatment planning system (TPS) as an IMU check for online adaptive radiotherapy using 1.5-Tesla MR-Linac.

View Article and Find Full Text PDF

Introduction: To investigate the impact of partial lateral scatter (LS), backscatter (BS) and presence of air gaps on optically stimulated luminescence dosimeter (OSLD) measurements in an acrylic miniphantom used for dosimetry audit on the 1.5 T magnetic resonance-linear accelerator (MR-linac) system.

Methods: The following irradiation geometries were investigated using OSLDs, A26 MR/A12 MR ion chamber (IC), and Monaco Monte Carlo system: (a) IC/OSLD in an acrylic miniphantom (partial LS, partial BS), (b) IC/OSLD in a miniphantom placed on a solid water (SW) stack at a depth of 1.

View Article and Find Full Text PDF

Purpose: To describe and report longitudinal quality assurance (QA) measurements for the mechanical and dosimetric performance of an Elekta Unity MR-linac during the first year of clinical use in our institution.

Materials And Methods: The mechanical and dosimetric performance of the MR-linac was evaluated with daily, weekly, monthly, and annual QA testing. The measurements monitor the size of the radiation isocenter, the MR-to-MV isocenter concordance, MLC and jaw position, the accuracy and reproducibility of step-and-shoot delivery, radiation output and beam profile constancy, and patient-specific QA for the first 50 treatments in our institution.

View Article and Find Full Text PDF

Objective: Adaptive radiation therapy (ART) is an integral part of MR-guided RT (MRgRT), requiring a new RT plan for each treatment fraction and resulting in a significant increase in patient-specific quality assurance (PSQA). This study investigates the possibility of using treatment log-file for automated PSQA.

Method: All treatment plans were delivered in 1.

View Article and Find Full Text PDF

Purpose/objective(s): The additional personnel and imaging procedures required for Adaptive Radiation Therapy (ART) pose a challenge for a broad implementation. We hypothesize that a change in transit fluence during the treatment course is correlated with the change of quality of life and thus can be used as a replanning trigger.

Materials/methods: Twenty-one head and neck cancer (HNC) patients filled out an MD Anderson Dysphagia Inventory (MDADI) questionnaire, before-and-after the radiotherapy treatment course.

View Article and Find Full Text PDF

Purpose: This study aimed to evaluate and compare different system calibration methods from a large cohort of systems to establish a commissioning procedure for surface-guided frameless cranial stereotactic radiosurgery (SRS) with intrafractional motion monitoring and gating. Using optical surface imaging (OSI) to guide non-coplanar SRS treatments, the determination of OSI couch-angle dependency, baseline drift, and gated-delivered-dose equivalency are essential.

Methods: Eleven trained physicists evaluated 17 OSI systems at nine clinical centers within our institution.

View Article and Find Full Text PDF

The use of trajectory log files for routine patient quality assurance is gaining acceptance. Such use requires the validation of the trajectory log itself. However, the accurate localization of a multileaf collimator (MLC) leaf while it is in motion remains a challenging task.

View Article and Find Full Text PDF

Over the last few years, magnetic resonance image-guided radiotherapy systems have been introduced into the clinic, allowing for daily online plan adaption. While quality assurance (QA) is similar to conventional radiotherapy systems, there is a need to introduce or modify measurement techniques. As yet, there is no consensus guidance on the QA equipment and test requirements for such systems.

View Article and Find Full Text PDF

Purpose: To evaluate the dosimetric effects of the AeroForm (AirXanpders®, Palo Alto, CA) tissue expander in-situ for breast cancer patients receiving post-mastectomy radiation therapy.

Methods And Materials: A film phantom (P1) was constructed by placing the metallic canister of the AeroForm on a solid water phantom with EBT3 films at five depths ranging from 2.6 mm to 66.

View Article and Find Full Text PDF

Background And Purpose: This study summarizes the cranial stereotactic radiosurgery (SRS) volumetric modulated arc therapy (VMAT) procedure at our institution.

Materials And Methods: Volumetric modulated arc therapy plans were generated for 40 patients with 188 lesions (range 2-8, median 5) in Eclipse and treated on a TrueBeam STx. Limitations of the custom beam model outside the central 2.

View Article and Find Full Text PDF

Purpose: To validate a machine learning approach to Virtual intensity-modulated radiation therapy (IMRT) quality assurance (QA) for accurately predicting gamma passing rates using different measurement approaches at different institutions.

Methods: A Virtual IMRT QA framework was previously developed using a machine learning algorithm based on 498 IMRT plans, in which QA measurements were performed using diode-array detectors and a 3%local/3 mm with 10% threshold at Institution 1. An independent set of 139 IMRT measurements from a different institution, Institution 2, with QA data based on portal dosimetry using the same gamma index, was used to test the mathematical framework.

View Article and Find Full Text PDF

Purpose: An ideal commissioning and quality assurance (QA) program for Volumetric Modulated Arc Therapy (VMAT) delivery systems should assess the performance of each individual dynamic component as a function of gantry angle. Procedures within such a program should also be time-efficient, independent of the delivery system and be sensitive to all types of errors. The purpose of this work is to develop a system for automated time-resolved commissioning and QA of VMAT control systems which meets these criteria.

View Article and Find Full Text PDF

Hypofractionated treatments generally increase the complexity of a treatment plan due to the more stringent constraints of normal tissues and target coverage. As a result, treatment plans contain more modulated MLC motions that may require extra efforts for accurate dose calculation. This study explores methods to minimize the differences between in-house dose calculation and actual delivery of hypofractionated volumetric-modulated arc therapy (VMAT), by focusing on arc approximation and tongue-and-groove (TG) modeling.

View Article and Find Full Text PDF