Purpose: To determine if patient-specific IMRT quality assurance can be measured on any matched treatment delivery system (TDS) for patient treatment delivery on another.
Methods: Three VMAT plans of varying complexity were created for each available energy for head and neck, SBRT lung, and right chestwall anatomical sites. Each plan was delivered on three matched Varian TrueBeam TDSs to the same Scandidos Delta Phantom+ diode array with only energy-specific device calibrations.
Background: Quality assurance measurement of IMRT/VMAT treatment plans is resource intensive, and other more efficient methods to achieve the same confidence are desirable.
Purpose: We aimed to analyze treatment plans in the context of the treatment planning systems that created them, in order to predict which ones will fail a standard quality assurance measurement. To do so, we sought to create a tool external to the treatment planning system that could analyze a set of MLC positions and provide information that could be used to calculate various evaluation metrics.
J Appl Clin Med Phys
August 2022
Purpose: We investigated the hypothesis that the transmission function of rounded end linearly traveling multileaf collimators (MLCs) is constant with position. This assumption is made by some MLC models used in clinical treatment planning systems (TPSs) and in the Varian MLC calibration convention. If not constant, this would have implications for treatment plan QA results.
View Article and Find Full Text PDFJ Appl Clin Med Phys
December 2021
A 6 MV flattened beam model for a Varian TrueBeamSTx c-arm treatment delivery system in RayStation, developed and validated at one institution, was implemented and validated at another institution. The only parameter value adjustments were to accommodate machine output at the second institution. Validation followed MPPG 5.
View Article and Find Full Text PDFJ Appl Clin Med Phys
June 2021
Purpose: To explore candidate RayStation beam models to serve as a class-specific template for a TrueBeam treatment delivery system.
Methods: Established validation techniques were used to evaluate three photon beam models: a clinically optimized model from the authors' institution, the built-in RayStation template, and a hybrid consisting of the RayStation template except substituting average MLC parameter values from a recent IROC survey. Comparisons were made for output factors, dose profiles from open fields, as well as representative VMAT test plans.
J Appl Clin Med Phys
March 2021
Purpose: To assess the implementation, accuracy, and validity of the dosimetric leaf gap correction (DLGC) in Mobius3D VMAT plan calculations.
Methods: The optimal Mobius3D DLGC was determined for both a TrueBeam with a Millennium multi-leaf collimator and a TrueBeamSTx with a high-definition multi-leaf collimator. By analyzing a broad series of seven VMAT plans and comparing the calculated to the measured dose delivered to a cylindrical phantom, optimal DLGC values were determined by minimizing the dose difference for both the collection of all plans, as well as for each plan individually.
The AAPM Medical Physics Practice Guideline (MPPG) 5.a provides concise guidance on the commissioning and QA of beam modeling and dose calculation in radiotherapy treatment planning systems. This work discusses the implementation of the validation testing recommended in MPPG 5.
View Article and Find Full Text PDFWe measured the X-ray transmission spectra of several biologically related samples in the phosphorus K-shell edge absorption region. These include red phosphorus, hydrated sodium phosphate (Na(3)PO(4).12 H(2)O), deoxyribonucleic acid (DNA), adenosine triphosphate (ATP), diolylphosphatidyl choline (DOPC), and Bacillus megaterium spores.
View Article and Find Full Text PDF