Background: Fragmentation, disconnection, or entrapment of an in-use microcatheter during neuro-endovascular procedures is a known risk. Often a benign entity, retained catheters are not infrequently observed, but severe complications including thrombus, thromboembolic events, pseudoaneurysm, and limb ischemia have been described, necessitating retrieval. This technical case report demonstrates the safe use of an external carotid artery (ECA) approach for ligation and removal of a retained microcatheter after middle meningeal artery (MMA) embolization.