Purpose: To develop and test an MRI coil assembly for imaging deep brain stimulation (DBS) at 3 T with a reduced level of local specific absorption rate of RF fields near the implant.

Methods: A mechanical rotatable linearly polarized birdcage transmitter outfitted with a 32-channel receive array was constructed. The coil performance and image quality were systematically evaluated using bench-level measurements and imaging performance tests, including SNR maps, array element noise correlation, and acceleration capabilities.

View Article and Find Full Text PDF

Purpose: To extend the coverage of brain coil arrays to the neck and cervical-spine region to enable combined head and neck imaging at 7 Tesla (T) ultra-high field MRI.

Methods: The coil array structures of a 64-channel receive coil and a 16-channel transmit coil were merged into one anatomically shaped close-fitting housing. Transmit characteristics were evaluated in a B -field mapping study and an electromagnetic model.

View Article and Find Full Text PDF

Purpose: Three 64-channel cardiac coils with different detector array configurations were designed and constructed to evaluate acceleration capabilities in simultaneous multislice (SMS) imaging for 3T cardiac MRI.

Methods: Three 64-channel coil array configurations obtained from a simulation-guided design approach were constructed and systematically evaluated regarding their encoding capabilities for accelerated SMS cardiac acquisitions at 3T. Array configuration A consists of uniformly distributed equally sized loops in an overlapped arrangement, B uses a gapped array design with symmetrically distributed equally sized loops, and C has non-uniform loop density and size, where smaller elements were centered over the heart and larger elements were placed surrounding the target region.