Publications by authors named "Saad Aldelaijan"

Purpose: To quantify and verify the dosimetric impact of high-dose rate (HDR) source positional uncertainty in brachytherapy, and to introduce a model for three-dimensional (3D) position tracking of the HDR source based on a two-dimensional (2D) measurement. This model has been utilized for the development of a comprehensive source quality assurance (QA) method using radiochromic film (RCF) dosimetry including assessment of different digitization uncertainties.

Methods: An algorithm was developed and verified to generate 2D dose maps of the mHDR-V2 Ir source (Elekta, Veenendaal, Netherlands) based on the AAPM TG-43 formalism.

View Article and Find Full Text PDF

The purpose of the present development is to employ 3D printing to prototype an ion chamber array with a scalable design potentially allowing increased spatial resolution and a larger active area. An additional goal is to design and fabricate a custom size thin-panel detector array with low-Z components. As a proof of principle demonstration, a medium size detector array with 30 × 30 air-vented ion chambers was 3D-printed using PLA as frame for the electrodes.

View Article and Find Full Text PDF

Purpose: To introduce a model that reproducibly linearizes the response from radiochromic film (RCF) dosimetry systems at extended dose range. To introduce a correction method, generated from the same scanned images, which corrects for scanner temporal response variation and scanner bed inhomogeneity.

Methods: Six calibration curves were established for different lot numbers of EBT3 GAFCHROMIC™ film model based on four EPSON scanners [10000XL (2 units), 11000XL, 12000XL] at three different centers.

View Article and Find Full Text PDF

Purpose: We investigate the effect of the GafChromic™ film EBT3 model absorbed dose energy response when used for dose measurements around low-energy photon sources. Monte Carlo based correction procedure in synergy with appropriate calibration curves was shown to provide more accurate absorbed dose (either relative or absolute). An assessment was made of possible dose errors that might be encountered if such energy dependent response is ignored.

View Article and Find Full Text PDF

Purpose: In this work we use Monte Carlo simulations to investigate change in Computed tomography (CT) X-ray energy spectra between exposures in air and within CT dose index (CTDI) phantom. While the results of these simulations will be relevant when measuring CTDI with any dosimeter, we apply the appropriate beam quality change correction for CTDI measurements using XR-QA2 model GafChromic™ film.

Methods: Dose profiles were measured with film strips, sandwiched between acrylic rods cut in half, placed within CTDI phantoms and scanned before and after irradiation with document scanner in reflective mode.

View Article and Find Full Text PDF

The primary source size is one of the most important beam model parameters in small photon fields. In this work we apply a recently suggested reconstruction technique to characterize the primary source of 6 Varian TrueBeam (TB) linacs. A series of photon fluence profile measurements were performed on 6 Varian TB linacs in the crossplane and inplane orientation using radiochromic film in air and a 2 mm Pb foil as a build-up layer.

View Article and Find Full Text PDF

Objective: Different dose response functions of EBT3 model GafChromic™ film dosimetry system have been compared in terms of sensitivity as well as uncertainty vs. error analysis. We also made an assessment of the necessity of scanning film pieces before and after irradiation.

View Article and Find Full Text PDF

Purpose: We compare image quality parameters derived from phantom images taken on three commercially available radiotherapy CT simulators. To make an unbiased evaluation, we assured images were obtained with the same surface dose measured using XR-QA2 model GafChromic™ film placed at the imaging phantom surface for all three CT-simulators.

Methods: Radiotherapy CT simulators GE LS 16, Philips Brilliance Big Bore, and Toshiba Aquilion LB were compared in terms of spatial resolution, low contrast detectability, image uniformity, and contrast to noise ratio using CATPHAN-504 phantom, scanned with Head and Pelvis protocols.

View Article and Find Full Text PDF

Purpose: To evaluate the response of the newest XR-QA2 GafChromic™ film model in terms of postexposure signal growth and energy response in comparison with the older XR-QA (Version 2) model.

Methods: Pieces of film were irradiated to air kerma in air values up to 12 cGy with several beam qualities (5.3-8.

View Article and Find Full Text PDF

Purpose: The Papillon technique using 50-kVp soft X-rays to treat rectal adenocarcinomas was developed and clinically implemented in the 1960s. We describe differences between accurate dosimetry and clinical implementation of this technique that is extending from its very inception to date.

Methods And Materials: A renaissance of the Papillon technique occurred with two recently introduced 50-kVp systems: Papillon+ by Ariane and a custom-made rectal applicator (consisting of a surface applicator inserted into a proctoscope) by iCAD's Xoft Axxent Electronic Brachytherapy (eBT) System (iCad, Inc.

View Article and Find Full Text PDF

Purpose: Current high-dose-rate brachytherapy skin treatments with the Freiburg flap (FF) applicator are planned with treatment planning systems based on the American Association of Physicists in Medicine TG-43 data sets, which assume full backscatter conditions in dose calculations. The aim of this work is to describe an experimental method based on radiochromic film dosimetry to evaluate dose calculation accuracy during surface treatments with the FF applicator at different depths and bolus thicknesses.

Methods And Materials: Absolute doses were measured using a reference EBT3 radiochromic film dosimetry system within a Solid Water phantom at different depths (0, 0.

View Article and Find Full Text PDF

Purpose: To describe an EBT3 GAFCHROMIC film-based dosimetry method to be used in commissioning of a combined HDR brachytherapy (HDRB) and stereotactic body radiation therapy (SBRT) boost for treatment of advanced cervical cancer involving extensive residual disease after external beam treatment.

Methods And Materials: A cube phantom was designed to firmly fit an intrauterine tandem applicator and EBT3 radiochromic film pieces. A high-risk clinical target volume (CTV) was contoured with an extended arm at one side.

View Article and Find Full Text PDF

Objective: Integration of fluorine-18 fludeoxyglucose ((18)F-FDG)-positron emission tomography (PET) functional data into conventional anatomically based gross tumour volume delineation may lead to optimization of dose to biological target volumes (BTV) in radiotherapy. We describe a method for defining tumour subvolumes using (18)F-FDG-PET data, based on the decomposition of differential uptake volume histograms (dUVHs).

Methods: For 27 patients with histopathologically proven non-small-cell lung carcinoma (NSCLC), background uptake values were sampled within the healthy lung contralateral to a tumour in those image slices containing tumour and then scaled by the ratio of mass densities between the healthy lung and tumour.

View Article and Find Full Text PDF

Purpose: Radiochromic films change their color upon irradiation due to polymerization of the sensitive component embedded within the sensitive layer. However, agents, other than monitored radiation, can lead to a change in the color of the sensitive layer (temperature, humidity, UV light) that can be considered as a background signal and can be removed from the actual measurement by using a control film piece. In this work, we investigate the impact of the use of control film pieces on both accuracy and uncertainty of dose measured using radiochromic film based reference dosimetry protocol.

View Article and Find Full Text PDF

Purpose: In the past, film dosimetry was developed into a powerful tool for external beam radiotherapy treatment verification and quality assurance. The objective of this work was the development and clinical testing of the EBT3 model GafChromic film based brachytherapy quality assurance (QA) system.

Methods And Materials: Retrospective dosimetry study was performed to test a patient-specific QA system for preoperative endorectal brachytherapy that uses a radiochromic film dosimetry system.

View Article and Find Full Text PDF

In this work we compare doses from imaging procedures performed on today's state-of-the-art integrated imaging systems using a reference radiochromic film dosimetry system. Skin dose and dose profile measurements from different imaging systems were performed using radiochromic films at different anatomical sites on a humanoid RANDO phantom. EBT3 film was used to measure imaging doses from a TomoTherapy MVCT system, while XRQA2 film was used for dose measurements from kilovoltage imaging systems (CBCT on 21eX and TrueBeam Varian linear accelerators and CyberKnife stereoscopic orthogonal imagers).

View Article and Find Full Text PDF

Purpose: The authors investigated the energy response of XR-QA2 GafChromic™ film over a broad energy range used in diagnostic radiology examinations. The authors also made an assessment of the most suitable functions for both reference and relative dose measurements.

Methods: Pieces of XR-QA2 film were irradiated to nine different values of air kerma in air, following reference calibration of a number of beam qualities ranging in HVLs from 0.

View Article and Find Full Text PDF

Purpose: Despite numerous advantages of radiochromic film dosimeter (high spatial resolution, near tissue equivalence, low energy dependence) to measure a relative dose distribution with film, one needs to first measure an absolute dose (following previously established reference dosimetry protocol) and then convert measured absolute dose values into relative doses. In this work, we present result of our efforts to obtain a functional form that would linearize the inherently nonlinear dose-response curve of the radiochromic film dosimetry system.

Methods: Functional form [ζ = (-1)[middle dot]netOD((2∕3))∕ln(netOD)] was derived from calibration curves of various previously established radiochromic film dosimetry systems.

View Article and Find Full Text PDF

Purpose: A radiochromic film based dosimetry system for high dose rate (HDR) Iridium-192 brachytherapy source was described. A comparison between calibration curves established in water and Solid Water™ was provided.

Methods: Pieces of EBT-2 model GAFCHROMIC™ film were irradiated in both water and Solid Water™ with HDR (192)Ir brachytherapy source in a dose range from 0 to 50 Gy.

View Article and Find Full Text PDF

Purpose: The authors present results of the measurements on the impact of radiochromic film immersion in water. The impact of film piece size, initial optical density, postimmersion waiting time prior to scanning, and the time film was kept in water has been investigated. The authors also investigated the pathways of water penetration into the film during the film immersion in water.

View Article and Find Full Text PDF

Purpose: One of the major drawbacks of the current radiochromic film dosimetry protocols is the postirradiation waiting time. In this article, the authors study the postirradiation time evolution of the absorption spectrum of radiochromic EBT-2 GAFCHROMIC film model.

Methods: Postirradiation scanning times range from 3 min to 5 days and a dose range extends from 0 to 6 Gy.

View Article and Find Full Text PDF