J Clin Monit Comput
February 1999
Objective: This study is an evaluation of a new thoracic bioimpedance cardiac output monitoring system which incorporates a modified form of the Kubicek equation and a method of estimating the left ventricular ejection time from the time derivative bioimpedance signals.
Methods: The performance of the new system was compared with conventional thermodilution in a porcine model. One hundred and ninety nine (n = 199) paired measurements of thermodilution cardiac output (TDCO) (range 1.
Objective: To evaluate the accuracy and precision of an advanced thoracic bioimpedance cardiac output monitor by comparing it with conventional thermodilution.
Design: Prospective data collected from 47 patients undergoing routine cardiac catheterization. The new bioimpedance system differs from its predecessors in electrode system configuration, advanced signal processing, use of a modified Kubicek equation, and a reliable estimate of left ventricular ejection time from the time derivative bioimpedance signals.