Infection with Toxoplasma gondii is often asymptomatic and, when acquired during pregnancy, may lead to connatal toxoplasmosis in the offspring. The newly introduced Vitros anti-Toxoplasma immunoglobulin G (IgG) and IgM assays, designed for the Vitros ECiQ immunodiagnostic system, a fully automated system based on chemiluminescence, were evaluated as a screening method for the serological detection of acute and chronic Toxoplasma infections in the sera of 719 pregnant women. The combination of the Vitros IgG and IgM assays demonstrated a sensitivity and a specificity of 100% for the successful detection of all acute T.
View Article and Find Full Text PDFThis is the first study comparing the caffeine testing by HPLC to the MicroTip Technology patented by Ortho-Clinical Diagnostics. For the determination of the precision for intra-run and day to day variances, control materials with concentration ranges between 2.3 microg/mL and 23.
View Article and Find Full Text PDFThe VITROS 5,1 FS analytical system was evaluated. The measurement of the serum proteins and HDL cholesterol with the newly developed MicroTip technology was compared with other nephelometric and turbidimetric methods. The within-run imprecision for apolipoprotein A1, apolipoprotein B, complement 3, complement 4, transferrin, immunoglobulin A, immunoglobulin M, immunoglobulin G and HDL cholesterol showed coefficients of variation between 0.