Publications by authors named "Ronnie K Binder"

Background: Computed tomography (CT) annular measurements are predictive of paravalvular regurgitation after transcatheter aortic valve replacement (TAVR) which is a predictor of mortality and morbidity.

Objectives: To demonstrate the rationale and potential implications of new CT sizing recommendations for TAVR.

Methods: The CT sizing recommendations aim to ensure routine transcatheter heart valve (THV) oversizing of the aortic annular area [(THV external area/systolic annular area - 1) × 100; range, 1%-20%; target, 10%-15%].

View Article and Find Full Text PDF