Purpose: To collect and analyze radiation dose-related data as part of international cooperation; to define diagnostic reference levels (DRL) for 24 X-ray projections in plain radiography (DX) considering anatomical region, clinical task, and procedural technique; and to harmonize the exposure practice across country borders.
Methods: A multicenter study was performed in Austria, Germany, Italy, and Switzerland in 2022-23 to provide dose-related data. Healthcare facilities were asked to provide processed data from their dose management systems.
The aim of this study is to propose diagnostic reference levels (DRLs) values for mammography in Switzerland. For the data collection, a survey was conducted among a sufficient number of centres, including five University hospitals, several cantonal hospitals, and large private clinics, covering all linguistic regions of Switzerland to be representative of the clinical practice. The data gathered contained the mean glandular dose (MGD), the compressed breast thickness (CBT), the mammography model and the examination parameters for each acquisition.
View Article and Find Full Text PDFObjective: To investigate the effect of two different large field of view (FOV) positions in the vertical dimension and shielding (thyroid collar and eyeglasses) on the effective dose and the local doses of various sites of the craniofacial complex.
Methods: Organ doses and effective doses were calculated based on the measured doses using 27 pairs of thermoluminescent dosemeters in a paediatric tissue-equivalent of a 10-year-old anthropomorphic phantom. The large FOV of the 3D Accuitomo F170 CBCT scanner was used to image parts of the craniofacial complex.