This paper describes the development of a simple, low cost chronoamperometric assay, for the measurement of fructose, using a graphite-nanoparticle modified screen-printed electrode (SPCE-G-COOH). Cyclic voltammetry showed that the response of the SPCE-G-COOH enhanced the sensitivity and precision, towards the enzymatically generated ferrocyanide species, over a plain SPCE; therefore the former was employed in subsequent studies. Calibration studies were carried out using chronoamperometry with a 40µl mixture containing fructose, mediator and FDH, deposited onto the SPCE-G-COOH.
View Article and Find Full Text PDFThis paper describes the optimisation of a screen-printing water-based carbon ink containing cobalt phthalocyanine (CoPC) and glucose oxidase (GOD) for the fabrication of a glucose biosensor. To optimise the performance of the biosensor, the loadings of the electrocatalyst (CoPC) and enzyme (GOD) were varied. It was found that the maximum linear range was achieved with a CoPC loading of 20% (m/m, relative to the mass of carbon) and a GOD loading of 628 U per gram of carbon.
View Article and Find Full Text PDFScreen-printed amperometric glucose biosensors have been fabricated using a water-based carbon ink. The enzyme glucose oxidase (GOD) and the electro-catalyst cobalt phthalocyanine were mixed with the carbon ink prior to the screen-printing process; therefore, biosensors are prepared in a one-step fabrication procedure. Optimisation of the biosensor performance was achieved by studying the effects of pH, buffer strength, and applied potential on the analytical response.
View Article and Find Full Text PDF