Purpose: Separate measurements are required when investigating multiple exercising muscles with singlevoxel-localized dynamic P-MRS. With multivoxel spectroscopy, P-MRS time-series spectra are acquired from multiple independent regions during one exercise-recovery experiment with the same time resolution as for singlevoxel measurements.
Methods: Multiple independently selected volumes were localized using temporally interleaved semi-LASER excitations at 7T.
Purpose: To enhance sensitivity and coverage for calf muscle studies, a novel, form-fitted, three-channel phosphorus-31 ((31) P), two-channel proton ((1) H) transceiver coil array for 7 T MR imaging and spectroscopy is presented.
Methods: Electromagnetic simulations employing individually generated voxel models were performed to design a coil array for studying nonpathological muscle metabolism. Static phase combinations of the coil elements' transmit fields were optimized based on homogeneity and efficiency for several voxel models.
Purpose: This article presents a novel inductive decoupling technique for form-fitting coil arrays of monolithic transmission line resonators, which target biomedical applications requiring high signal-to-noise ratio over a large field of view to image anatomical structures varying in size and shape from patient to patient.
Methods: Individual transmission line resonator elements are mutually decoupled using magnetic flux sharing by overlapping annexes. This decoupling technique was evaluated by electromagnetic simulations and bench measurements for two- and four-element arrays, comparing single- and double-gap transmission line resonator designs, combined either with a basic capacitive matching scheme or inductive pickup loop matching.