Aim: During radiotherapy, in patients with implantable cardioverter-defibrillators (ICDs) malfunctions are considered more likely if doses more than 2 Gy reach the ICD site; however, most malfunctions occur with high-energy (>10 MV) radiations, and the risk is less defined using 6-MV linear accelerators. The purpose of the study is to experimentally evaluate the occurrence of malfunctions in ICDs radiated with a 6-MV linear accelerator at increasing photon doses.
Methods: Thirty-two ICDs from all manufacturers (31 explanted and one demo) were evaluated; all devices with a sufficient battery charge underwent multiple radiations with a 6-MV photon beam reaching a cumulative dose at ICD site of 0.
Progress in biomaterial technology and improvements in surgical and perfusion strategy ameliorated morbidity and mortality in pediatric cardiac surgery. In this study, we describe our clinical experience comparing performance of two neonatal oxygenators. From January 2002 to March 2011, 159 infants with less than 5 kg body weight underwent heart surgery.