The purpose of the present work is to evaluate the feasibility of a novel real-time beam monitoring device for medical linacs which remotely senses charge carriers produced in air by the beam without intersecting and attenuating the beamline. The primary goal is to elaborate a theoretical concept of a possible detector geometry and underlying physical model that allows for determination of clinically relevant beam data in real time, namely MLC leaf positions and dose rate. The detector consists of two opposing electrode arrays arranged in two possible orientations around the beamline.