Publications by authors named "Rie Muramatsu"

Purpose: The spot position is an important beam parameter in the quality assurance of scanning proton therapy. In this study, we investigated dosimetric impact of systematic 15 spot position errors (SSPE) in spot scanning proton therapy using three types of optimization methods of head and neck tumor.

Materials And Methods: The planning simulation was performed with ± 2 mm model SSPE in the X and Y directions.

View Article and Find Full Text PDF

Purpose: Accurate calculation of the proton beam range inside a patient is an important topic in proton therapy. In recent times, a computed tomography (CT) image reconstruction algorithm was developed for treatment planning to reduce the impact of the variation of the CT number with changes in imaging conditions. In this study, we investigated the usefulness of this new reconstruction algorithm (DirectDensity™: DD) in proton therapy based on its comparison with filtered back projection (FBP).

View Article and Find Full Text PDF

The purpose of this study was to evaluate the effect of quality assurance (QA)-related setup errors in passive proton therapy for prostate cancer with and without a hydrogel spacer. We used 20 typical computed tomography (CT) images of prostate cancer: 10 patients with and 10 patients without spacers. The following 12 model errors were assumed: output error ± 2%, range error ± 1 mm, setup error ± 1 mm for three directions, and multileaf collimator (MLC) position error ± 1 mm.

View Article and Find Full Text PDF

The purpose of this study was to provide periodic quality assurance (QA) methods for respiratory-gated proton beam with a range modulation wheel (RMW) and to clarify the characteristics and long-term stability of the respiratory-gated proton beam. A two-dimensional detector array and a solid water phantom were used to measure absolute dose, spread-out Bragg peak (SOBP) width and proton range for monthly QA. SOBP width and proton range were measured using an oblique incidence beam to the lateral side of a solid water phantom and compared between with and without a gating proton beam.

View Article and Find Full Text PDF

Purpose: Suppression of respiratory movement of the liver would be desirable for high-precision radiation therapy for liver tumors. We aimed to investigate the effect of our original device-free compressed shell fixation method and breathing instruction on suppression of respiratory movement. The characteristics of liver motion based on the movement of a fiducial marker were also analyzed.

View Article and Find Full Text PDF

In this study, we evaluate dosimetric advantages of using patient-specific aperture system with intensity-modulated proton therapy (IMPT) for head and neck tumors at the shallow depth. We used four types of patient-specific aperture system (PSAS) to irradiate shallow regions less than 4 g/cm with a sharp lateral penumbra. Ten head and neck IMPT plans with or without aperture were optimized separately with the same 95% prescription dose and same dose constraint for organs at risk (OARs).

View Article and Find Full Text PDF