Background: ZAP-70 protein expression has been proposed as a marker for immunoglobulin heavy chain mutational status, which some studies have correlated with disease course in B-cell chronic lymphocytic leukemia (CLL). Studies published to date measuring levels of expression of ZAP-70 intracellular protein using flow cytometry have demonstrated poor performance, as defined by the difference in signal in known positive and negative lymphocyte populations.
Methods: A recently published method (Chow S, Hedley DW, Grom P, Magari R, Jacobberger JW, Shankey TV, Cytometry A 2005;67:4-17) to measure intracellular phospho-epitopes was optimized using a design of experiments (DOE) approach to provide the best separation of ZAP-70 expression in positive T- or NK-cells as compared to negative B-cells in peripheral blood samples.