In this work, the apparent treatment dose that kV planar or CBCT imaging contributes to Gafchromic EBT3 film used for in vivo dosimetry, was investigated. Gafchromic EBT3 film pieces were attached to a variety of phantoms and irradiated using the linear accelerator's built-in kV imaging system, in both kV planar mode and CBCT mode. To evaluate the sensitivity of the film in the clinical scenario where dose contributions are received from both imaging and treatment, additional pieces of film were irradiated using base doses of 50cGy and then irradiated using selected kV planar and CBCT techniques.
View Article and Find Full Text PDFAustralas Phys Eng Sci Med
September 2013
Radiochromic film has the potential to provide accurate in vivo dosimetry measurements. However, it is not known whether small film pieces can still provide accurate dosimetric results. The use of small film pieces is of particular interest in regions of interest (ROIs) such as the eye, or where the patient's contour changes rapidly.
View Article and Find Full Text PDF