Sporadic mechanically-induced hemolysis associated with kinks in extracorporeal blood circuits during hemodialysis is a rare but potentially serious complication that exhibits laboratory features consistent with both in vivo and in vitro hemolysis. Misclassification of clinically significant hemolysis as in vitro can lead to inappropriate test cancellation and delayed medical interventions. Here, we report three cases of hemolysis attributed to kinked hemodialysis blood lines, which we have defined as "ex vivo" hemolysis.
View Article and Find Full Text PDF