Publications by authors named "Rafael Yankelevich"

The purpose of this study is to investigate (1) the depth at which the measurement of the block transmission factor should be made, and (2) the level of the transmission of 18 and 22 MeV electron beams through conventional Cerrobend. We measured the block transmission in water phantom as ionization profiles across the beam and as ionization distributions along the central axis of the beam for 18 and 22 MeV electron beams, for cone sizes ranging from 6 x 10 cm2 to 25 x 25 cm2. In our analysis, we separated the bremsstrahlung component produced in the Cerrobend block from the component originating in the head in the transmitted dose under the standard Cerrobend block.

View Article and Find Full Text PDF

Contemporary radiation oncology departments are often lacking a conventional simulator due to common use of virtual simulation and recent implementation of image guided radiation therapy. A protocol based on MammoSite method was developed using CT based planning, a Source Position Simulator (SPS) with a Simulator Wire and a linear accelerator based On-Board Imager (OBI) for daily verification. After MammoSite balloon implantation, the patient undergoes a CT study.

View Article and Find Full Text PDF