Publications by authors named "Radu Racasan"

There have been a number of reports on the occurrence of taper corrosion and/or fretting and some have speculated on a link to the occurrence of adverse local tissue reaction specifically in relation to total hip replacement which have a metal-on-metal bearing. As such a study was carried out to compare the magnitude of material loss at the taper in a series of retrieved femoral heads used in metal-on-polyethylene bearings with that in a series of retrieved heads used in metal-on-metal bearings. A total of 36 metal-on-polyethylene and 21 metal-on-metal femoral components were included in the study all of which were received from a customer complaint database.

View Article and Find Full Text PDF

It has been speculated that material loss, either as corrosion or wear, at the head-stem taper junction is implicated in the high revision rates reported for metal-on-metal total hip replacements. We measured the volume of material loss from the taper and bearing surfaces of retrieved devices, and investigated the associations with blood metal ion levels and the diagnosis of a cystic or solid pseudotumor. The median volumes of material lost from the female and male taper surfaces were 2.

View Article and Find Full Text PDF