As a commonly used solution, the multi-ended readout can measure the depth-of-interaction (DOI) for positron emission tomography (PET) detectors. In the present study, the effects of the multi-ended readout design were investigated using the leading-edge discriminator (LED) triggers on the timing performance of time-of-flight (TOF) PET detectors. At the very first, the photon transmission model of the four detectors, namely, single-ended readout, dual-ended readout, side dual-ended readout, and triple-ended readout, was established in Tracepro.
View Article and Find Full Text PDFPreclinical positron emission tomography (PET) is a sensitive and quantitative molecule imaging modality widely used in characterizing the biological processes and diseases in small animals. The purpose of this study is to investigate the methods to optimize a PET detector for high-resolution preclinical imaging. The PET detector proposed in this study consists of a 28 × 28 array of LYSO crystals 0.
View Article and Find Full Text PDFPortable radiation detectors are widely used in environmental radiation detection and medical imaging due to their portability feature, high detection efficiency, and large field of view. Lutetium-yttrium oxyorthosilicate (LYSO) is a widely used scintillator in gamma radiation detection. However, the structure and the arrangement of scintillators limit the sensitivity and detection accuracy of these radiation detectors.