Annu Int Conf IEEE Eng Med Biol Soc
March 2011
An online algorithm for determining respiratory mechanics in patients using non-invasive ventilation (NIV) in pressure support mode was developed and embedded in a ventilator system. Based on multiple linear regression (MLR) of respiratory data, the algorithm was tested on a patient bench model under conditions with and without leak and simulating a variety of mechanics. Bland-Altman analysis indicates reliable measures of compliance across the clinical range of interest (± 11-18% limits of agreement).
View Article and Find Full Text PDFAnnu Int Conf IEEE Eng Med Biol Soc
March 2010
An automated classification algorithm for the detection of expiratory ineffective efforts in patient-ventilator interaction is developed and validated. Using this algorithm, 5624 breaths from 23 patients in a pulmonary ward were examined. The participants (N = 23) underwent both conventional and non-invasive ventilation.
View Article and Find Full Text PDFObjective: Imperfect patient-ventilator interaction is common during assisted ventilation, and the detection of clinically relevant mismatching requires visual monitoring of the ventilator screen. We have assessed the feasibility, sensitivity and specificity of an algorithm embedded in a ventilator system that is able to automatically detect the occurrence of ineffective triggering and double triggering in real time.
Design: Prospective study.