In this study, the hypothesis was tested that a low-resistant, high-compliant oxygenator provides better pulse conductance and less hemolysis than a high-resistant, low-compliant oxygenator during pulsatile cardiopulmonary bypass. Forty adults undergoing coronary artery bypass surgery were randomly divided into two groups using either an oxygenator with a relatively low hydraulic resistance (Quadrox BE-HMO 2000, Maquet Cardiopulmonary AG, Hirrlingen, Germany) or with a relatively high hydraulic resistance (Capiox SX18, Terumo Cardiovascular Systems, Tokyo, Japan). The phase shift between the flow signals measured at the inlet and outlet of the oxygenator was used to assess compliance.
View Article and Find Full Text PDFVenous cannulae undergo continuous improvements to achieve better and safer venous drainage. Several cannula tests have been reported, though cannula performance during inlet obstruction has never been a test criterion. In this study, five different cannulae for proximal venous drainage were tested in a mock circulation that enabled measurement of hydraulic conductance after inlet obstruction by vessel collapse.
View Article and Find Full Text PDF