Purpose: An advantage of rebound tonometry (RT) is its ease of use so that it can also be operated by health care technicians. However, the cost of the disposable measuring probes is high and their reuse carries the risk of infection. Therefore, this study aims to objectify the potential risk of bacterial transmission by RT.
View Article and Find Full Text PDFGraefes Arch Clin Exp Ophthalmol
November 2017
Purpose: For diaphanoscopy or transscleral laser applications, the transmission of the sclera is an essential property. The study aimed to determine the pressure dependent transmission of human sclera from anterior to posterior.
Methods: Pressure dependent transmission measurements were performed by a pressure inducing setup at the range of 60-2058 kPa.
Existing light sources for intraocular illumination are often bulky and expensive and pose a risk for the patient, because light guides are inserted in the eye through incisions and if the tip of these light guides get too close to the retina, the retina can be damaged photochemically within minutes or even seconds. Therefore a new, safe and simple device for intraocular illumination is developed and evaluated for its thermal and photochemical risks to the patient. It consists of a white LED which is integrated into a pen like holder.
View Article and Find Full Text PDF