Purpose: Preclinical studies using a new treatment modality called FLASH Radiotherapy (FLASH-RT) need a two-phase procedure to ensure minimal uncertainties in the delivered dose. The first phase requires a new investigation of the reference dosimetry lying outside the conventional metrology framework from national metrology institutes but necessary to obtain traceability, repeatability, and stability of irradiations. The second consists of performing special quality assurance procedure prior to irradiation.
View Article and Find Full Text PDFPurpose: The Oriatron eRT6 is an experimental high dose-per-pulse linear accelerator (linac) which was designed to deliver an electron beam with variable dose-rates, ranging from a few Gy/min up to hundreds of Gy/s. It was built to study the radiobiological effects of high dose-per-pulse/dose-rate electron beam irradiation, in the context of preclinical and cognitive studies. In this work, we report on the commissioning and beam monitoring of the Oriatron eRT6 prototype linac.
View Article and Find Full Text PDF