Publications by authors named "Philipe Pibarot"

Background: Reference values of hemodynamic parameters for the assessment of prosthetic heart valves are necessary, and ideally need to be provided by entities independent of the valve manufacturers. Thus, the study aim was to provide, in vitro, normal reference values of the effective orifice area (EOA) for different models and sizes of mitral prosthetic valve, and to assess the determinants of EOA and mean transvalvular pressure gradient (mTPG).

Methods: Four models of mechanical prostheses were tested (one mono-leaflet, three bi-leaflet) and four models of bioprostheses (two bovine pericardial, two porcine) on a double-activation pulsed duplicator that was specifically designed and optimized for assessing the hemodynamic performance of mitral prosthetic valves.

View Article and Find Full Text PDF

Unlabelled: Rubidium-ARMI ((82)Rb as an Alternative Radiopharmaceutical for Myocardial Imaging) is a multicenter trial to evaluate the accuracy, outcomes, and cost-effectiveness of low-dose (82)Rb perfusion imaging using 3-dimensional (3D) PET/CT technology. Standardized imaging protocols are essential to ensure consistent interpretation.

Methods: Cardiac phantom qualifying scans were obtained at 7 recruiting centers.

View Article and Find Full Text PDF